Electric Force (EFM) Imaging
Electric Field Gradient Detection—Procedures
Rev. B MultiMode SPM Instruction Manual 257
3. Mount a metal-coated NanoProbe cantilever into the electric field cantilever holder. MFM-
style cantilevers (225µm long, with resonant frequencies around 70kHz) usually work well.
It is also possible to deposit custom coatings on model FESP silicon TappingMode
cantilevers. Make sure that any deposited metal you use adheres strongly to the silicon
cantilever.
4. Set up the AFM as usual for TappingMode operation. In the Channel panels, be certain all
Highpass and Lowpass filters are Off. Set the Rounding parameter in the Microscope >
Calibrate > Scanner window to zero (0.00).
5. Select View > Cantilever Tune.
6. Follow the instructions below for the type of electric force imaging desired, Phase Detection
or Amplitude Detection.
14.4.1 Phase Detection
Phase Detection is only available when the Basic Extender Module has been correctly congured
into the system.
In the Cantilever Tune window, set Start frequency and End frequency to appropriate
values for your cantilever (e.g., for 225µm MFM cantilevers, set Start frequency to
40kHz and End frequency to 100kHz). Select Autotune.
•Two curves appear on the Cantilever Tune graph: the amplitude curve in white, and the
phase curve in yellow. (In Figure 14.4e, the phase curve is the dashed line and the
amplitude curve is the solid line).
Figure 14.4b Phase detection Cantilever Tune
(for systems with the Basic Extender Module installed)
The phase should decrease with increasing frequency and cross the center line (90° point) at the
peak frequency. The phase curve then correctly reects the phase lag between the drive voltage and
the cantilever response. Again, gradients in the electric force will cause a shift F
0
in the resonance
frequency. In this case, resonance shifts give rise to phase shifts φ which can then give an image of
the electric force gradients; see Figure 14.4c.
Seitenansicht 276
1 2 ... 272 273 274 275 276 277 278 279 280 281 282 ... 357 358

Kommentare zu diesen Handbüchern

Keine Kommentare

Dimplex SCx175si manuels

Manuels d'utilisation et guides de l'utilisateur pour Chauffe-eau et chaudières Dimplex SCx175si.
Nous fournissons des manuels en pdf 1 Dimplex SCx175si à télécharger gratuitement par type de document : Spécifications






Plus de produits et de manuels pour Chauffe-eau et chaudières Dimplex

Modèles Type de document
WB5E 1 Instructions d'exploitation   Dimplex WB5E 1 Operating instructions, 3 pages
AWP 300 LW Manuel de service   Dimplex AWP 300 LW Technical data, 20 pages
DEE 2102 Manuel de l'utilisateur      Montageanweisung DEC 1802 / 2102 / 2402, 22 pages
SCxn150d Manuel d'installation   Dimplex SCxn150d Installation manual, 6 pages
System 18 Instructions d'exploitation   Dimplex System 18 Operating instructions, 8 pages
WB7 Manuel d'instructions   Dimplex WB7 Instruction manual, 13 pages